We unpack the FDA 510(k) summary for Tyto Insights for Wheeze Detection (with PCCP). Listeners will learn how this over-the-counter AI software analyzes lung sounds, the real performance numbers behind it, and exactly how the FDA's Predetermined Change Control Plan allows algorithms to be retrained post-clearance without new submissions.

Key points

  • The software is an over-the-counter AI tool that analyzes 6 to 12-second acoustic lung signals from a compatible stethoscope to detect suspected wheezing in patients 2 years and older.
  • Performance testing showed a specificity of 0.99 and a sensitivity of 0.55, meaning it rarely flags false positives but misses a significant portion of actual wheezes.
  • This 510(k) clearance is based on substantial equivalence to a 2023 version of the same product, with the changes being cybersecurity updates and the addition of a Predetermined Change Control Plan (PCCP).
  • The PCCP allows the company to retrain the machine learning model to improve performance without filing a new 510(k), provided the new version's sensitivity and specificity do not drop below the currently validated thresholds.
  • The FDA approval letter emphasizes that even with a PCCP, all software modifications must still follow strict Quality Management System Regulations, including design controls and documentation in the Medical Device File.

Source: FDA 510(k) summary K252844: Tyto Insights for Wheeze Detection (with PCCP) (Tyto Care) - U.S. Food and Drug Administration, 2026